PET-NHP

Very Large bore PET module for PET, PET-CT and PET-MRI Studies

Exceptional PET Upgrade for the compatible MR Solutions series of MRI’s and CT’s for sequential PET-MRI and PET-CT studies of large animals and NHP

  • Key Features

    • Benchtop design Can be detached from the MRI or CT for independent PET imaging
    • Sequential PET-MRI imaging at 3.0 Tesla
    • Sequential PET-CT with CT-XL
    • Can be operated independently to the MRI or CT
    • Enables Multi-modality imaging

The PET-NHP is designed for large animal and NHP imaging studies
The PET-NHP has an inner bore size of 311 mm and provides an active trans-axial field of view of 220mm.
The system offers the latest technology with true Depth of Interaction (DOI). This enables resolutions of below 1,1mm across the whole field of view to be
achieved.
The PET-NHP can be combined with the CT-NHP or the 3T-XL  MRI (large-bore cryogen-free system with 40cm outer diameter).

Clear-Bore:                                 311 mm

Transaxial FOV:                        220 mm

Axial FOV:                                 206.64 mm

Extended FOV (with bed):     400 mm

Dual Lyso DOI dtectors

Spatial Resolution:                  1.1 mm (3DOSEM)